Title: Extending the range of amperometric sensors
Other Titles: Розширення діапазону вимірювань амперометричних сенсорів
Authors: Buket, Oleksandr
Linyucheva, Olga
Nahorniy, Oleksiy
Bludenko, Andriy
Linyuchev, Oleksandr
Bibliographic description (Ukraine): Extending the range of amperometric sensors / Oleksandr Buket, Olga Linyucheva, Oleksiy Nahorniy, Andriy Bludenko, Oleksandr Linyuchev // Chemistry & Chemical Technology. – 2015. – Volume 9, number 2. – P. 251–255. – Bibliography: 9 titles.
Issue Date: 2015
Publisher: Publishing House of Lviv Polytechnic National University
Keywords: chlorine sensor
amperometric type
measurement range
upper limit
electrolyte pH
сенсор хлору
амперметричний тип
діапазон вимірювань
верхня межа
рН електроліту
Abstract: It was experimentally and theoretically shown that the upper range limit of a chlorine sensor is limited by the component of chlorine mass transfer in the liquid phase of the electrolyte solution. To extend the range, it was proposed to enhance absorption of chlorine by the electrolyte through increasing its pH. Приведено теоретичні та експериментальні докази, що верхня межа діапазону вимірювань сенсора хлору обмежується складовою масопереносу хлору в рідкій фазі електроліту. Для розширення діапазону вимірювань запропоновано збільшити абсорбцію хлору електролітом внаслідок підвищення його рН.
URI: https://ena.lpnu.ua/handle/ntb/28149
Content type: Article
Appears in Collections:Chemistry & Chemical Technology. – 2015. – Vol. 9, No. 2

Files in This Item:
File Description SizeFormat 
021_123_128.pdf192.14 kBAdobe PDFView/Open